EasyManua.ls Logo

Dräger Fabius plus - Calibrating the flow sensor

Dräger Fabius plus
232 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Fabius plus SW 3.n 133
Configuration
The test results (B) are displayed on the screen.
After completion of the system test, the total test
results (C) are displayed, see chapter ''Checking
the readiness for operation'' on page 88. If the
system test was successful, the Standby mode is
activated.
Calibrating the flow sensor
1 Press the Calibrate Flow Sensor softkey (A).
2 Follow the instructions on the screen.
At the start of the calibration, the instructions are
hidden and the following message is displayed
above the standby softkeys (B):
Flow Calibration in progress
After the calibration, one of the following two
messages are displayed above the standby
softkeys (B):
Flow Calibration completed - reconnect
expiratory hose
or
Flow Calibration Failed
Troubleshooting with failed flow calibration
Repeat the calibration.
Replace the flow sensor.
If the calibration continues to fail, contact
DrägerService or the authorized local service
partner.
2549425463
B
C
A
B
Standby

Table of Contents

Other manuals for Dräger Fabius plus

Related product manuals