Assembly and preparation
80 Instructions for use Fabius plus SW 3.n
Connection of the pressure gauge for
measurement of the airway pressure
(optional)
1 Connect the pressure measurement hose (C)
to the corresponding connection on the bottom
of the compact breathing system.
2 Connect the T-piece (B) with the pressure
measurement hose and the pressure gauge.
3 Connect the other end of the pressure
measurement hose with the bacterial filter (B).
4 Plug in the bacterial filter into the connection
(D) marked with on the rear of the device.
Connecting the flow sensor
1 Connect the flow sensor cable to the
corresponding connection (A) on the bottom of
the compact breathing system.
2 Plug in the other end of the flow sensor cable
into the connection (B) marked with on the
rear of the device.
25447
A
B
C
D
25449
A
B