Switching the volume alarms on and off
1 Press the Volume Alarms ON/OFF softkey
(A).
The key label changes from Volume Alarms
ON to Volume Alarms OFF.
Instead of the upper and lower alarm limits, the
symbol appears, indicating the alarm is
deactivated.
The volume alarms are deactivated.
Automatic setting of the pressure
threshold
1 Press the Auto Set softkey (A).
The pressure threshold for the peak pressure
(PEAK) is set to 4 cmH
2O (hPa) below the
current plateau pressure (PLAT).
25505
NOTE
The function Volume Alarms ON /OFF is
available in the standard view of the ManSpont
mode. If the key is pressed in ManSpont
mode, the softkey Volume Alarms ON /OFF is
not displayed.
25505
NOTE
The pressure threshold must not be below
5cmH
2O (5 hPa) or above 30 cmH2O (30 hPa).
NOTE
If no current measured value is available for the
plateau pressure (PLAT), pressing the softkey
has no effect.
NOTE
In SIMV/PS mode, the pressure threshold
depends on the pressure of the mandatory
breaths.