Instructions for use Fabius plus SW 3.n 123
Monitoring
Parameter field for O2 monitoring
The following information is displayed in the
parameter field for O
2 monitoring:
A Numerical value for the inspiratory O
2
concentration in percent (%) in the range from
10 % to 100 %
B Upper alarm limit for the O
2 concentration in
(%)
C Lower alarm limit for the O
2 concentration in
(%)
Setting the O2 alarm limits
The standard alarm limits configured for the
ventilation mode can be used unchanged, see
chapter ''Restoring the factory settings'' on page
142.
or
The alarm limits can be set individually for the
current case:
1 Press the (A) key.
The dialog window with the alarm limits (B) opens.
2 Adjust the upper and lower alarm limit values of
the O
2 concentration (C), see setting ranges of
the alarm parameters in chapter ''Changing the
alarm limits'' on page 141.
3 Confirm new values.