Instructions for use Fabius plus SW 3.n 171
Cleaning, disinfection and sterilization
Information concerning dismounted
accessory parts and attached devices
Observe the instructions for use of the following
accessory parts:
Accessory parts
Flow sensor
CLIC adapter
CLIC absorber, Infinity ID CLIC absorber
Breathing hoses
Filter
Breathing bag
–Masks
Water trap
Vaporizer
Attached devices
Endotracheal suction
Hinged arms
–Monitors
Sensors and cables
IT systems
–AGS
Single-use articles without instructions for use
Soda lime dust filter (optional)
Sample line
WARNING
Risk of infection
Used sample lines may be infectious due to
the breathing gases that passed through
them.
Replace the sample lines regularly, see table
''Semicritical medical devices''.
CAUTION
Material damage due to disinfectants
When the sample line is disinfected and residues
of the agent remain in the sample line, these
residues can get into the water trap and the gas
measurement module later. This may result in
faulty measurements.
Sample lines are single-use items and may not
be disinfected.