Technical data
208 Instructions for use Fabius plus SW 3.n
Essential performance characteristics
The essential performance features comprise:
– Supplying the anesthesia workstation with O
2
If the O2 supply (central gas supply or gas
cylinder) fails, an alarm is issued.
– Supply of the patient with adequately
oxygenated breathing gas
If the breathing gas contains insufficient levels
of O
2, an alarm is issued.
– Monitoring of the airway pressure and the
expiratory minute volume
Alarms are issued depending on the set alarm
limits.
– Measurement accuracy of the O
2
measurement.
Alarms are issued depending on the set alarm
limits. If the O
2 sensor fails, an alarm is issued.
Pin assignment
Pin 1 n/c
Pin 2 TXD
Pin 3 RXD
Pin 4 n/c
Pin 5 GND
Pin 6 n/c
Pin 7 n/c
Pin 8 n/c
Pin 9 n/c
NOTE
In accordance with general safety standards,
additional components are required for a
complete anesthesia workstation.