Instructions for use Fabius plus SW 3.n 149
Configuration
Calibrating the O2 sensor
1 Press the Calibrate O2 Sensor softkey (A).
2 Follow the instructions on the screen.
To calibrate the O2 sensor, proceed as described
in chapter ''Calibrating the O2 sensor'' on page
134.
During the calibration, the O
2 value in the window
(B) of the O2 monitoring is replaced by the word
CAL. The calibration time is approx. 15 seconds.
After successful calibration, the O
2 measured
value is again displayed.
If the calibration was not successful, replace
the O
2 sensor capsule in the O2 sensor
housing, see chapter ''Inserting a new O
2
sensor capsule'' on page 78.
If the calibration continues to fail, contact
DrägerService or the authorized local service
partner.