Instructions for use Fabius plus SW 3.n 47
Operating concept
Fresh-gas delivery (version for 3 gases)
The flow tubes and the pressure gauges are
located on the front of the device below the screen.
There are 2 flow control valves for Air and O
2 on
the gas mixer for 2 gases. There is a third flow
control valve for N
2O on the gas mixer for 3 gases.
All flow control valves are labeled and provided
with a color coding, see chapter ''Color coding for
anesthetic agents and medical gases'' on page 49.
The O2 flow control valve is additionally provided
with grooves on the gripping surface.
A Pressure gauge for the central O
2 supply
B Pressure gauge for the central Air supply
C Pressure gauge for the central N
2O supply
D Flow tubes
E O
2 flow control valve
F Air flow control valve
G N
2O flow control valve
If the pressure drops below 20 psi (1.4 kPa x100),
the LED warning indicator lights up.
25384
NOTE
If the float does not turn, the accuracy may be
less than that specified in the technical data.