Cleaning, disinfection and sterilization
180 Instructions for use Fabius plus SW 3.n
Semicritical medical devices
Items which can be
reprocessed
Recommended
reprocessing
intervals
Pre-
cleaning
Machine
cleaning and
disinfection
Manual Steriliza-
tion
Cleaning Disinfec-
tion
Breathing system:
– Breathing system
housing
– Inspiratory/expira-
tory ports, APL
valve
– Inspiratory valve,
expiratory valve
– Bag elbow
– Rigid arm for
breathing bag
(optional)
– Flexible arm for
breathing bag
(optional)
Weekly Yes Yes Possible Possible Possible
– Breathing hoses After each patient Observe the associated instructions for use.
Absorber container
and absorber insert
Weekly Yes Yes Possible Possible Possible
Soda lime dust filter
(optional)
Replace each
time soda lime is
changed.
No No No No No
Sample line Replacement
only
No No No No No
– When the sample
line is fitted to the
filter on the Y-
piece.
Daily
– When the sample
line is fitted directly
to the Y-piece and
the filters are fit-
ted on the breath-
ing system
After each patient
Ventilator lid After each patient Yes Yes No Yes Yes
Ventilator membrane After each patient Yes Yes No Yes Yes
Ventilator hose After each patient Observe the associated instructions for use.