Instructions for use Fabius plus SW 3.n 179
Cleaning, disinfection and sterilization
Other surfaces which are fre-
quently touched:
– Side parts of the housings of
the screen and of other patient
monitors
– Accessory parts:
– Storage trays
– Shelf
– Hinged arms
– Probes of compressed gas
hoses
– Mains plug
– Gas cylinder valves
– Transfer hose of the anes-
thetic gas receiving system
– Cables and hoses that lie
on floor
–Brake
Daily Outside Outside
Surfaces which are touched less
frequently:
– Network cables and data
cables
– Compressed gas hoses
– Pressure reducers
– Gas cylinders
– Drawer surfaces, outside and
inside
– Anesthetic gas receiving sys-
tem
– Holder for sample line
Weekly Outside Outside
Items which can be reprocessed Recommended repro-
cessing intervals
Manual
Cleaning Disinfection