132 Instructions for use Fabius plus SW 3.n
If the flow control valves are not closed before
accessing Standby mode, the following message
(D) is shown on the start screen:
Gas still flowing - shut off all gas flow control
valves to activate Sleep Mode.
As soon as the flow control valves are closed, the
message disappears.
3 Close the flow control valves.
Power-saving mode
When Fabius is in Standby mode and there is no
user input for 2.5 minutes, the power-saving mode
is activated. The screen is then replaced by the
screensaver. Press any key to end the
screensaver.
Performing the system test
In Standby mode, a system test can be started.
This test corresponds to the test that is performed
automatically after switching on the anesthesia
workstation. More information can be found in
chapter ''Switching on'' on page 86.
1 Press the Run System Test softkey (A).
Electrical system components are tested.
Default settings are restored.
2549225493
D
Standby
WARNING
Risk of patient injury
During the system test, the system is
pressurized.
To prevent patient injury, do not perform the
system test on the medical device if a patient
is connected.
25463
A