Calibrating the O2 sensor
In order for the O2 sensor to be correctly
calibrated, it must be exposed to the ambient air
during the entire calibration.
To avoid leakage, remove the O
2 sensor from the
cover of the inspiratory valve. Seal the cover of the
inspiratory valve with the valve cover plug.
1 Press the Calibrate O
2 Sensor softkey (A).
2 Follow the instructions on the screen.
At the start of the calibration, the instructions are
hidden and the following message is displayed
above the standby softkeys (B):
O
2 Calibration in progress
After the calibration, one of the following two
messages are displayed above the standby
softkeys (B):
O
2 Sensor Calibration completed - reinsert O2
sensor
or
O
2 Sensor Calibration Failed
Troubleshooting with failed O
2 calibration
Replace the O
2 sensor capsule in the O2
sensor housing, see chapter ''Inserting a new
O
2 sensor capsule'' on page 78.
If the calibration continues to fail, contact
DrägerService or the authorized local service
partner.