Monitoring
122 Instructions for use Fabius plus SW 3.n
Main screen
Press the (A) key.
The current dialog window changes to the main
screen.
The following information is displayed on the
screen:
Current alarm messages
Data of the O
2 monitoring
Data of the airway pressure monitoring
Data of the breathing volume monitoring
O2 monitoring
The inspiratory oxygen concentration is measured
with a dual galvanic sensor, which is located in the
cover of the inspiratory valve, see chapter
''Compact breathing system COSY (top view)'' on
page 22.
25411
A
CAUTION
Risk of inaccurate measured values
If the O
2 sensor is removed, this can lead to
leaks in the breathing system.
When the O2 sensor is replaced or removed, it
must be recalibrated.
NOTE
If the anesthesia workstation is not used, remove
the O
2 sensor from the cover of the inspiratory
valve and insert the sealing plug provided.