Technical data
200 Instructions for use Fabius plus SW 3.n
Anesthetic gas supply module
Fresh-gas delivery, low-flow measuring tubes
(calibrated in accordance with IEC 60601-2-13 for
ambient air 20 °C (68 °F) (±3 °C), 1013 cmH
2O
(hPa):
Range and accuracy:
O
2 0.02 to 10.0 L/min ±10 % of the indicated value
for flows between 1 and 10 L/min
N
2O 0.02 to 10.0 L/min ±10 % of the indicated value
for flows between 1 and 10 L/min
Air 0.2 to 12.0 L/min ±10 % of the indicated value for
flows between 1 and 12 L/min
Fresh-gas flow stability
O
2 and N2O: ±10% of the setting at supply pressures of 41 to 87 psi (2.8 to 6 kPa x 100)
Air: ±10% of the setting at supply pressures of 50 to 55 psi (3.4 to 3.8 kPa x 100).
Outside the range of 50 to 55 psi (3.4 to 3.8 kPa x 100), the Air flow varies proportionally to the supply
pressures.
O
2 flush at 87 psi (6 kPa x 100): max. 75 L/min
at 41 psi (2.8 kPa x 100): min. 25 L/min
Pressure limit of the common gas outlet Maximum 13 psi (0.9 kPa x 100) ±5%
Flow tube for O
2 supplemental delivery (optional)
Connection Stepped connection for use with hoses of different
diameters
Flow 0 to 10 L/min
Accuracy ±5 % of the measurement range
Resolution 0.5 L/min