Instructions for use Fabius plus SW 3.n 161
Troubleshooting
Alarm
priority
Alarm Cause Remedy
(31) AIRWAY PRESSURE HIGH !!! The upper alarm limit for the
airway pressure was
exceeded, the breathing hose
is kinked.
Check the breathing circuit
connected to the anesthe-
sia workstation.
The alarm limit was set too
low.
Check the breathing sys-
tem or the alarm limit.
(23/31) APNOEA FLOW !! The apnea flow alarm based
on a time staggering.
In the Volume Control,
Pressure Control, SIMV/PS
modes with Freq 6 or in
Pressure Support mode
with apnea ventilation deacti-
vated:
Caution = VT <20 mL for >15
seconds
In the ManSpont, SIMV/PS
modes with Freq < 6 or in
Pressure Support mode
with apnea ventilation acti-
vated:
Caution = VT <20 mL for >30
seconds
Breathing/ventilation
stopped.
Check ventilator.
Leakage or disconnection in
the breathing system.
Check breathing system.