Instructions for use Fabius plus SW 3.n 163
Troubleshooting
(23/31) APNOEA PRESSURE !! The apnea pressure alarm is
based on a time staggering.
In the Volume Control,
Pressure Control, SIMV/PS
modes with Freq 6 or in
Pressure Support mode
with apnea ventilation deacti-
vated:
Caution = PAW did not
exceed the pressure thresh-
old value for a duration of >15
seconds.
In the ManSpont, SIMV/PS
modes with Freq < 6 or in
Pressure Support mode
with apnea ventilation acti-
vated:
Caution = PAW did not
exceed the pressure thresh-
old value for a duration of >30
seconds.
Breathing/ventilation
stopped.
Check ventilator.
Leakage or disconnection in
the breathing system.
Check breathing system.