Instructions for use Fabius plus SW 3.n 193
Technical data
Opening pressure of the internal safety valve 70 psi (4.8 kPa x 100)
Protection class
Device I, in compliance with IEC 60601-1
Application parts, connections for breathing hoses Type BF
Classification in compliance with Directive
93/42/EEC, Annex IX
II b
UMDNS Code Universal Medical Device Nomen-
clature System - nomenclature system for medi-
cal devices
10-134
GMDN Code Global Medical Device Nomencla-
ture - worldwide nomenclature for medical
devices
37710
Use of latex Not made with natural rubber latex.
Penetration of liquids IP20 in accordance with IEC 60529
Power supply
The power rating cannot be configured, with optional
Dräger power socket strip (see instructions for use of
power socket strip for medical devices, 9038776)
100 to 240 VAC, 50/60 Hz, 2.3 A max
Internal battery
Power rating 24 V; 3.5 Ah
Type Closed, lead/acid, gel
Charge time 16 hours on mains power for full operating time
Backup time with fully charged battery Minimum 45 minutes