44 Instructions for use | Oxylog VE300 SW 1.n
Assembly and preparation
Connecting the breathing hose and the flow measuring lines
1. Connect the breathing hose (A) to the breathing valve.
2. Connect the flow measuring lines (B) to the ports of the flow sensor. When
connecting the flow measuring lines, pay attention to the differing diameters of
the hoses and connectors and connect them to the correct connector.
3. Connect the flow measuring lines to the device. Correct alignment is indicated
by a groove on the connector, which must point away from the breathing hose. If
it is incorrectly seated, incorrect values will be displayed.
4. Connect the breathing hose to the device's gas outlet.
5.7.3 Connecting the disposable breathing circuit for adults
Using disposable hoses can reduce the risk of cross-infection. Disposable hoses
are shipped clean but non-sterile.
WARNING
Risk of patient injury
Soiling of the disposable breathing circuit may put the patient at risk.
► Open the packaging of disposable breathing circuits only immediately before
use.
1. Connect the pressure measuring line to the device.
2. Connect the breathing hose to the device's gas outlet.
5.7.4 Connecting the disposable breathing circuit for adults
(Plus option)
Using disposable hoses can reduce the risk of cross-infection. Disposable hoses
are shipped clean but non-sterile.
WARNING
Risk of patient injury
Soiling of the disposable breathing circuit may put the patient at risk.
► Open the packaging of disposable breathing circuits only immediately before
use.
1. Connect the flow measuring lines to the device. Correct alignment is indicated
by a groove on the connector, which must point away from the breathing hose. If
it is incorrectly seated, incorrect values will be displayed.
2. Connect the breathing hose to the device's gas outlet.
30066