Instructions for use | Oxylog VE300 SW 1.n 43
Assembly and preparation
5.7.2 Connecting the reusable breathing circuit for adults
(Plus option)
Assembly of the breathing valve
WARNING
Risk of CO2 rebreathing
An incorrectly assembled or malfunctioning breathing valve may put the patient at
risk.
► Do not remove, damage, or bend the rubber disk (A) in the housing.
1. Place the diaphragm (B) in the breathing valve housing (C). Make sure that it is
fitted correctly.
2. Mount the cover (A) and turn it about 60° clockwise until it engages.
3. Push the flow sensor (D) onto the breathing valve (C). To do so, align the flow
sensor (D) on the breathing valve (C) using the grooves in the flow sensor.
4. Push the elbow (E) onto the flow sensor (D).
3006430065