417101774_EMP_KKS_E60_E.doc - 57 - Rev. 8-03.10
10.4.4 Calibration / Manual
If the calibra
tion value to be entered is known, a “dry calibration” (immediate entry of the
value without a previous calibration process) can be conducted.
However, this method is not very accurate as no consideration is given to the on-site
conditions (back-pressure, viscosity, cable cross-chapters and lengths etc).
Fig. 10.59
10.4.4.1 Calibration data table
The pump capacity in l/h
is allocated to the pump’s electronics on the basis of the calibration
values entered in the table.
Example:
A specified value of 23.3 ml results in a pump capacity of 1.4 l/h.
NOTE
These values relate to the metering medium of water at 20°C
Pump Pump capacity [l/h] Pump calibration value [ml]
0,2 3,33
0,5 8,33
0,9 15,0
EMP KKS
1,4 23,3
10.5 Operating data
The following operating data is recorded and displayed under this menu item:
Operating hours
Liter
Amount of plulses