EasyManua.ls Logo

Ecotech Serinus 44 - Instrument Gain and Cell Pressure Issues

Ecotech Serinus 44
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Serinus 44 User Manual 3.0
Page 130
Error Message/Problem
Cause
Solution
Error Message/Problem
Cause
Solution
Flow block temp failure
Faulty heater or
temperature sensor
Refer to Section 7.4.
Reaction cell temp failure
Faulty heater or
temperature sensor
Refer to Section 7.5.
Instrument resetting
Multiple possibilities
1. Check that the instrument is not overheating.
2. Possibly a faulty power supply.
3. Corrupted firmware. Perform ‘erase all settings’ in
the Bootloader Menu and reload or upgrade
firmware.
12 Voltage supply failure
Power supply has failed
Replace power supply.
No display
AC power
Verify that the mains power cable is connected and
the rear fan is operating
Contrast misadjusted
Adjust the display contrast by pressing two keys on
the front panel:
Press up arrow () for darker contrast
Press down arrow () for lighter contrast
DC power
Verify that the power supply is providing ±12V +5V
DC. TP40, TP23 and TP41
Display
Check the interface cable between the display and
the microprocessor board.
Bad display or
microprocessor PCA
1. Replace the front panel display.
2. Replace the microprocessor board.
3. A bad cable is unlikely, but if you suspect it,
perform a pin-for-pin continuity test using an
ohmmeter.
Sample pressure
too high or too low
Loss of pressure
calibration
Ensure particulate filter has been recently changed.
Ensure tubing is not kinked or blocked. Ensure
vacuum pump is correctly installed and operating.
Perform a pressure calibration (refer to section 5.2).
Sample flow not at 0.9
SLPM
Multiple possibilities
Check/replace sample filter.
Check pump.
Check valves.
Check/replace sintered filter.
Re-calibrate pressure sensors.
Unstable flow or pressure
readings
Faulty pressure sensors
Check pressure transducer calibration. Check
calibration valve block is functioning and not
blocked. If unable to diagnose problem check voltage
across TP1 and TP2 of manifold pressure PCA ensure
it is about 4 volts ±0.5 volts. Check voltage across
TP1 and TP2 of cell pressure PCA (with vacuum
connected) ensure it is about 1 volts ±0.5 volts. If
unable to diagnose problem, it may be a noisy A/D
converter, replace main controller PCA.

Table of Contents

Related product manuals