A leak in the instrument or calibration system dilutes
the sample stream and causes low span readings and
noise.
Adjust the span using the calibration procedure.
Leaks or blockages in tubes or valves. Perform leak
check and flow check and repair any leaks/blockages.
Faulty calibration source
Ensure calibration gas is plumbed correctly, is not
contaminated, has no leaks and is a certified
reference gas.
Faulty optical bench or ozone generator.
Not enough zero air
supplied
Check calibrator excess flow is sufficient.
Ensure zero air sources are not overly polluted.
Lower than expected NO
X
and NO
2
NH
3
scrubber, NO
2
converter
Preform a NO
2
converter efficiency test (refer to
Section 5.7).
Lower than expected N
X
and NH
3
NH
3
converter, NH
3
selective scrubber, NO
2
converter
Preform a NH
3
converter efficiency test (refer to
Section 5.8).