61
Gas 1 Offset
A offset applied to Gas 1
62
Gas 3 Offset
A offset applied to Gas 3
63
Flow Temperature
Temperature of the flow block
64
Lamp Current
Lamp current in mA E.g. 35mA
65
Digital Supply
Digital Supply voltage (should always read close to 5
volts)
66
Conc. Voltage
Concentration Voltage
67
High Voltage
High Voltage reading for PMT
68
Ozonator
0=Off, 1=On
69
Control Loop
0=Off, 1=On (default is On)
70
Diagnostic Mode
0=Operate
1=Preamp
2=Electrical
3=Optical
(default is Operate)
71
Gas Flow
72
Gas Pressure
73
Ambient Pressure
74
12V Supply
The 12 volt Power supply voltage
75
Cell Temperature
Cell Temperature
76
Conv. Temperature
Converter Temperature
77
Chassis Temperature
Chassis Temperature
78
Manifold Temp.
Manifold Temperature
79
Cooler Temperature
Cooler Temperature
80
Mirror Temperature
Mirror Temperature
81
Lamp Temperature
Lamp Temperature
82
O3 Gen Lamp Temperature
O3 Lamp Temperature
83
Instrument Status
84
Reference Voltage
85
Calibration State
0 = MEASURE
1 = CYCLE
2 = ZERO
3 = SPAN
86
Primary Raw Conc.
(For S40, before NOx background and gain)
87
Secondary Raw Conc.
Only for multigas instruments
(For S40, before NOx background and gain)
88
S40 Backgnd Conc.
NOx Background Concentration
(For S40, before gain)