Serinus 44 User Manual 3.0
Page 164
173
Gas3 Over Range
The Upper Concentration Range when Over-Ranging is
enabled for Analog Output 3
174
Gas1 Over-Ranging
0 = Over Ranging Disabled
1 = Over Ranging Enabled (Gas1)
175
Gas2 Over-Ranging
0 = Over Ranging Disabled
1 = Over Ranging Enabled (Gas2)
176
Gas3 Over-Ranging
0 = Over Ranging Disabled
1 = Over Ranging Enabled (Gas3)
177
Heater Set Point
Cell Heater Set Point
178
Reserved
179
PMT Test LED Pot
PMT Test LED intensity controller POT
180
Last Power Failure
Time Stamp of the Last power fail (4 byte time stamp)
Bit 31:26 ---- Year (0 99)
Bit 25:22 ---- Month ( 1 12)
Bit 21:17 ---- Date (1 31)
Bit 16:12 ---- Hour (00 23)
Bit 11:06 ---- Min (00 59)
Bit 05:00 ---- Sec (00 59)
181
Inst Manifld Pres.
Instantaneous Manifold Pressure in S40 instrument (no
filter)
182
Cell Press. (gas3)
E.g. NH3 in an S44
183
Gas 4 Conc.
E.g. NH3 in an S44
184
Gas 4 Avg.
185
Gas 5 Conc.
E.g. NH3 on an S44
186
NH3 Conv.Efficiency
Gas 3 (eg:NO2)
187
Cell/Lamp Duty Cycle
188
Mirror T. Duty Cycle
189
Flow Temp Duty Cycle
190
Cooler T. Duty Cycle
191
Conv Temp Duty Cycle
192
CO Conv T Duty Cycle
193
F/Scale Curr Gas 1
194
F/Scale Curr Gas 2
195
F/Scale Curr Gas 3
196
Z Adj Curr Gas 1
197
Z Adj Curr Gas 2
198
Z Adj Curr Gas 3
199
Ext Analog Input 1

Table of Contents