Parts List and Schematics
Page 165
200
Ext Analog Input 2
201
Ext Analog Input 3
202
H2S Conv Set Point
Converter Set Point
203
Cal. Pressure 3
Calibration Pressure 3
204
PTF Correction (gas 3)
Pressure Temperature Flow Compensation Factor for
third gas in multigas instruments.
205
Dilution Ratio
The current dilution ratio (default is 1.0)
206
Traffic Light
State of the status light:
0 = Green
1 = Amber
2 = Off (normally impossible)
3 = Red
207
Network Protocol
0 = EC9800
1 = Bayern-Hassen
2 = Advanced
3 = Modbus
208
Gas 4 Offset
209
O3 GEN Fine Pot
Ozone generator control. Output value depends on
whether the instrument is pot or DAC controlled.
Pot: 0..255
DAC: 0..64535
210
O3 Gen Lamp Current
mA
211
O3 GEN Coarse Pot
Ozone generator control
Pot: 0..99
DAC: Repeat of parameter 209
212
Logging Period
The data logging period, in seconds (1.. 86400)
213
O3 Gen Coeff A
Ozone generator coefficients
214
O3 Gen Coeff B
215
O3 Gen Coeff C

Table of Contents