Ecotech Serinus 44 - Appendix B. EC9800 Protocol; EC9800 Protocol Commands

Ecotech Serinus 44
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Serinus 44 User Manual 3.0
Page 166
Appendix B. EC9800 Protocol
The following commands are supported.
DCONC
Function: Sends the current instantaneous concentration data to the serial port.
Format: DCONC,{<DEVICE I.D.>}{TERMINATOR}
Device response: {GAS}<SPACE>{STATUS WORD}<CR><LF>
All numbers are in floating point format. The STATUS WORD indicates the instrument status in hex
using the following format:
Bit 15 = SYSFAIL (MSB)
Bit 14 = FLOWFAIL
Bit 13 = LAMPFAIL
Bit 12 = CHOPFAIL
Bit 11 = CVFAIL
Bit 10 = COOLERFAIL
Bit 9 = HEATERFAIL
Bit 8 = REFFAIL
Bit 7 = PS-FAIL
Bit 6 = HV-FAIL
Bit 5 = OUT OF SERVICE
Bit 4 = Instrument is in zero mode
Bit 3 = Instrument is in span mode
Bit 2 = Unused
Bit 1 = SETPPM selected, CLEARMG/M3
Bit 0 = reserved (LSB).
DSPAN
Function: Commands the unit under test to enter the span mode and stay there.
Format: DSPAN,{<DEVICE I.D.>}{TERMINATOR}
Device response: <ACK> if the unit under test is able to perform the command, <NAK> if not
DZERO
Function: Commands the unit under test to enter the zero mode and stay there.
Format: DZERO,{<DEVICE I.D.>}{TERMINATOR}
Device response: <ACK> if the unit under test is able to perform the command, <NAK> if not
ABORT
Function: Commands the addressed device to abort the current mode and return to
the measure mode.
Format: ABORT,{<DEVICE I.D.>}{TERMINATOR}
Device response: <ACK> if the unit under test is able to perform the command, <NAK> if not