SCU-800 Control Unit for Turbomolecular Pump
*1
The driver output stops and the rotor continues rotating by inertia.
It may take several hours to stop. To stop the STP pump quickly, close the vacuum valve at the outlet port
flange and introduce gas from the purge port into th
e STP pump.
Table
35 - Error
messages after self test (continued)
1. When changing the cable
length, perform tuning.
2. Contact
Service office
.
1. Remove external vibration
so as not to transmit it to
the STP
pump.
2. Check
the vacuum piping.
3. Install the STP
pump in a
way that no foreign matter
falls into the STP
pump.
1. Contact
Service office
.
2. Contact
Service office
.
1. Contact
Service office
.
2. Contact
Serv
ice office
.
Abnormal magnetic bearing:
1. Change of the cable
length.
2. Disconnection of the STP
connection cable or control
circuit error.
Excessive vibration applied
externally to the STP
pump:
1. External vibration/impact.
2. Atmospheric air fl
ows into
the STP
pump.
3. Foreign matter falls into
the STP
pump.
1. Wear
-out of the cooling
water bearing
.
2. Failure of the cooling fan
drive circuit
.
1. Short of the STP
connection
cable.
Contact noise and
vibration of the rotor
are generated from
the STP
pump.
Decelerate and stop.
Failure of serial
communication.
Abnormal driver operation
.
Cooling fan come to a
stop
.
Disturbance X_H
Disturbance Y_B
Disturbance Y_H
Disturbance Z