SCU-800 Control Unit for Turbomolecular Pump
*1
The driver output stops and the rotor continues rotating by inertia.
It may take several hours to stop. To stop the STP pump quickly,
close the vacuum valve at the outlet port
flange and introduce gas from the purge port into the STP pump.
Table
36 - Error
messages after self test (continued)
1. Use the maximum
working pressure or less.
2. Use the allo
wable backing
pressure or less.
3. Check the vacuum piping
system for leakage.
4. Check the backing
-pump
for its capacity and
START state
.
(use a backing
-pump with
the recommended
capacity or more)
1. Connect the cable
correctly. If the cable is
not locked, lock it
securely.
2. Set "
Leak
Valve Option"
to "DISABLE".
1. High pressure at the inlet
port.
2. High pressure at the outlet
port.
3. Leakage of the piping
system.
4. Failure of the
backing
-pump.
1. Disconnection of the
emergency vent valve
cable.
2. Incorrect setting of the
emergency vent valve.
Touch down. The
magnetic bearing
output stops and the
rotor is descended
on the
touch down
bearing.
ACCELERATION state
continues for
approximately 30 min
utes.
Failure of the driver circuit
.
Serial communication
failure.
Disconnection of the
emergency valve cable.