Emerson Process Management GmbH & Co. OHG2-14
X-STREAM X2FD
Instruction Manual
HASXMDE-IM-EX
01/2015
2.5 Installation - Electrical
The next page shows
the English installation instructions for type
AD-U adapters, reprinted with the permis-
sion of
Redapt Ltd
Note!
See the documentation delivered together
with your analyzer for an original version of
the cable gland or adapter installation instruc-
tions, whatever is applicable !
INSTALLATION USING CABLE GLANDS
X-STREAM Flameproof analyzers are shipped without certi ed cable glands.
Cable glands have to be selected while preparing the installation according
all applicable standards, e. g. EN 60079-14 or IEC 60079-14.
Use only cables as speci ed in the cable glands installation instruction,
and carefully follow the original manufacturer‘s installation instructions,
accompanying the cable glands!