Emerson Process Management GmbH & Co. OHGS-6
X-STREAM X2FD
Instruction Manual
HASXMDE-IM-EX
01/2015
Safety Instructions
Authorized Personnel
In-depth specialist knowledge is an absolutely necessary condition for working with and on the
analyzer.
Authorized personnel for installing, operating, servicing and maintaining the analyzer are instructed
and trained quali ed personnel of the operating company and the manufacturer.
It is the responsibility of the operating company to
train staff,
observe safety regulations,
follow the instruction manual.
Operators must
have been trained,
have read and understood all relevant sections of the instruction manual before commencing
work,
know the safety mechanisms and regulations.
To avoid personal injury and loss of property, do not install, operate, maintain or service this instru-
ment before reading and understanding this instruction manual and receiving appropriate training.
Special Conditions for Safe Use
• Only speci ed screws M16x45 ISO 4762
A2-70 as speci ed in the maintenance
section of this manual shall be used (spare
part # 42716945).
• The ame joints correspond to the drawing
No. 4.271-7112/1 and do not comply with
the dimensions mentioned into the Tab. 1
and Tab. 2 of EN 60079-1 ed.2.
The gas path for the sample gas shall be
equipped with additional appropriate ame
arrestors in case of gas pressure above
1100 hPa to 1500 hPa.
• Appropriate certi ed cable glands shall be
used in accordance with IEC/EN 60079-14.
Vapor recovery application:
Pressure of gases not to exceed 1100 hPa.
Concentrations of gases must be below
25 % LEL.
Depending on the particular application all
approbriate safety instructions mentioned
in this instruction manual on hand must
be considered!
Take special care of formation of amma-
ble gas at the outlet of breathing and/or
purging devices, if the sample gas con-
centration is above 25% LEL! If need be,
such outlets have to end in a safe area!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson ROSEMOUNT Xstream X2 Series and is the answer not in the manual?

Emerson ROSEMOUNT Xstream X2 Series Specifications

General IconGeneral
BrandEmerson
ModelROSEMOUNT Xstream X2 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals