Emerson Process Management GmbH & Co. OHG1-4
X-STREAM X2FD
Instruction Manual
HASXMDE-IM-EX
01/2015
1.5 Compliances
1.5 Explosion Protection Compliances
This product is available in two different varia-
tions, separately certi ed by agencies for the
use in hazardous (classi ed) areas:
The one version, to be equipped with cable
glands, is certi ed by Fyzikálně technický
zkušební ústav, s.p (FTZÚ), an European
Noti ed Body under the Directive 94/9/EC
(„ATEX“) and conforms to the provisions of
EN 60079-0 and EN 60079-1. See appen-
dix for a copy of the EC type examination
certi cate.
The second variation, to be equipped with
metric-to-NPT adapters and conduits (these
components are not part of the instrument
certi cation), is certi ed by the Canadian
Standards Association, an „OSHA Na-
tionally Recognized Testing Laboratory“
(NRTL), for Canada and USA and conforms
to the provisions of CAN/CSA-E60079-0:02
(R2006), CAN/CSA-E60079-1:02 (R2006),
ANSI/ISA-12.00.01-2002 (IEC 60079-0 Mod),
ANSI/ISA-12.22.01-2002 (IEC 60079-1 Mod).
Furthermore, these X-STREAM X2FD analy-
zers are certi ed for use in Class I, Division 2,
Group BCD T3 areas.
See appendix for a copy of the CSA Certi cate
of Compliance.
IECEx certi cation enables worldwide appro-
vals with minimized need of testing. Further-
more, based on this approval several national
approvals are granted.
For a comprehensive list of applicable certi-
cates visit our website at
www.emersonprocess.com.
The following certi cation markings apply to
the products:
European Union (EU, ATEX)
Category 2, Zone 1:
Ex d IIB+H
2
T4 Gb
EC ATEX Type Examination Certi cate:
FTZU 08 ATEX 0028 X.
IECEx
Ex d IIB+H
2
T4 Gb
Conforms to the provisions of the „Equipment
intended for use in Potentially Explosive At-
mospheres (ATEX)“ Directive 94/9/EC, EMC
Directive 2004/108/EC and CE Directive
93/68/EEC.
USA
Class I, Zone 1, AEx d IIB+H
2
T3
Class I, Division 2, Group BCD T3
Canada
Class I, Zone 1, Ex d IIB+H
2
T3
Certi cate of Compliance 1714037X

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson ROSEMOUNT Xstream X2 Series and is the answer not in the manual?

Emerson ROSEMOUNT Xstream X2 Series Specifications

General IconGeneral
BrandEmerson
ModelROSEMOUNT Xstream X2 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals