1.5 Compliances
1.5.1 Special conditions for safe use
• Only specifi ed screws M16x45 ISO 4762
A2-70 as specifi ed in the maintenance
section of this manual shall be used (spare
part # 42716945).
• The fl ame joints correspond to the drawing
No. 4.271-7112/1 and do not comply with
the dimensions mentioned into the Tab. 1
and Tab. 2 of EN 60079-1 ed.2.
• The gas path for the sample gas shall be
equipped with additional appropriate fl ame
arrestors in case of gas pressure above
1100 hPa to 1500 hPa.
• Appropriate certifi ed cable glands shall be
used in accordance with IEC/EN 60079-14.
• Vapor Recovery application:
Pressure of gases not to exceed 1100 hPa.
Concentrations of gases must be below
25 % LEL.
• Depending on the particular application all
approbriate safety instructions mentioned
in this instruction manual on hand must
be considered!
• Take special care of formation of fl amma-
ble gas at the outlet of breathing and/or
purging devices, if the sample gas con-
centration is above 25% LEL! If need be,
such outlets have to end in a safe area!
EAC Certifi cate Russia
1 Ex d IIB + H2 T4 X
Certifi cate no.:ТС RU C-DE.ГБ04.В.00327
Approved by: CTB
KGS Korea
Ex d IIB+H2 T4
Certifi cate no.: 13-GA4BO-0649X
Approved by KGS