Emerson Process Management GmbH & Co. OHG3-26
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
3.6 Measurement Specications
All performance data are veried during the manufacturing process for each unit by the following
tests:
Linearization and sensitivity test
Long term drift stability test
Climate chamber test
Cross interference test (if applicable)
Note 1!
Not all data listed are applicable to all analyzer versions (e.g. 60 °C thermostatically controlled
box is not available for electrochemical and trace oxygen, nor for
1
219 in instruments).
Note 2!
For NDIR/UV/VIS measurements, take into account that
sample gas may diffuse or be released by leakages into the analyzer enclosure
if existent in the analyzer surroundings, the component to be measured may enter the enclosure.
Concentrations then may increase inside the enclosure. High concentrations of the component
to be measured inside the enclosure may inuence the measurement by unintended absorption,
which could cause drift of the measurement.
A remedy for this issue is to purge the housing with gas not containing the component of interest.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals