Emerson Process Management GmbH & Co. OHG 5-17
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
5
Startup
Signal damping (set in DAMPING) allows
smoothing the output signal, but also affects
the response time of outputs and display:
The factory setting is 0 seconds.
The maximum possible t
90
time is limited by
the size of the internal sampling buffer and
the sampling rates of the installed measu-
ring principles/sensors.
Multi-channel units: the value for each
channel must be entered separately.
The second page´s last line
enables the user
to enter the current ambient pressure ma-
nually, if no pressure sensor is installed,
or
to view the current pressure, if a sensor
is installed ( page 5-14).
If no pressure sensor is installed, enter the
current ambient pressure here and adjust
it, when signicant changes take place: this
improves the accuracy of the instrument.
Press
left several times to return to SETUP
and open MEASUREMENT.
5.7 Checking the Settings
Setup..
Measurement..
Ch1
Measurement 1of2
Ranges..
Damping..
Linearization..
Cross Interference..
Average..
Delay..
Ch1
Measurement 2of2
Cut-off: None
Pressure Compensation ..
Ch1
Pressure compensation
Manual Pressure: 1013 hPa
Pressure 1013 hPa
Pressure Status Good

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals