Maintenance & Procedures
7.4.5 Unattended Automatic Calibration
All submenus opened by above menu lines
have the same content, exemplarily described
below:
Set to Enabled, to use interval times for the
selected calibration (here: Zero all)
Enter the interval time
Accepted range: 1 ... 10,000 h
In this submenu enter the start time for the
rst calibration (s. below)
In this submenu enter the date and start
time for the rst calibration after nishing this
setup.
Note!
Time format is 24h (1 pm = 13)
Press
left to return to the previous menu, to
see a summary.
Start time & date for the next calibration,
based on the entered parameters (empty
until date & time have been entered in the
submenu)
Start time & date for the next calibration,
based on the entered parameters
Current date & time
...
Enable: Enabled
Interval: 15 h
Start time..
Next: --
Time 10/01/10 15:33
...
Enable: Enabled
Interval: 15 h
Start time..
Next: 10/01/10 16:00
Time 10/01/10 15:33
Start...
Month: 1
Day: 10
Hour: 16
Minute: 0
Set!
Next 10/01/10 16:00
Note!
If the displayed current time is not correct,
update the system setup on
6-100.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals