Emerson Process Management GmbH & Co. OHG7-48
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Ch1
IF Source1
Source: None
Value 0
Status Good
Interference Factor
: 1
Adjust IF Factor!
Linearization Curve..
8. After all data has been entered, press left
to return to the previous menu.
To automatically calculate the proper
interference compensation factor, enter
„Adjust IF Factor!“ and press enter.
The analyzer now starts calculating and,
when nished, shows the result in line
„Interference Factor:“
Note!
If it turns out during measurement, that
the calculated factor is not satisfactory,
you may either change the „Interference
Factor“ manually (simply enter the line,
press enter and change the value), or
adjust the polynomials (repeat step 7).
7.5 Cross Interference Compensation
Don‘t forget to set „Activated“
to the previous value after -
nishing the cross interference
compensation setup!
If no additional sources of interference are
to be congured, press home to return to the
MEASUREMENT SCREEN.
9. If need be,
• repeat from step 4 for more sources of
interference for the currently selected
channel,
or
• start from step 1 to congure interference
compensation for another channel.