Emerson Process Management GmbH & Co. OHG8-4
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Message
Status level
Explanation Recommended Actions
E-Mail: Not sent
Maintenance
Failed to send e-mail
Check Ethernet connection
Check SMTP conguration
E-Mail: Could not open
LOG le
Maintenance
Accessing log le not possib-
le
No NAMUR!
Check if internal disk is pre-
sent / installed
Call Emerson service
External Failure
Failure
An external source (e.g. digital
input or PLC program) meets
a failure condition that is for-
warded to the self-diagnostics
system.
Check the assigned digital
input for the condition.
Check PLC program for the
condition.
Reassign digital inputs to not
being forwarded to the dia-
gnostic system.
External FctCheck
Function check
An external source (e.g.
digital input or PLC program)
meets a function check con-
dition that is forwarded to the
self-diagnostics system.
External MaintRequ
Maintenance
An external source (e.g. digital
input or PLC program) meets
a maintenance request con-
dition that is forwarded to the
self-diagnostics system.
External OffSpec
Off-spec
An external source (e.g.
digital input or PLC program)
meets a out-of-specication
condition that is forwarded to
the self-diagnostics system.
Factory le open error
Maintenance
Opening the factory congura-
tion le is not possible.
Create a new factory con-
guration le
Check the le system for
consistency using CHKDISK
tool
8.2.1 Analyzer Related Messages