Emerson Process Management GmbH & Co. OHG 8-17
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
8
Troubleshooting
Situation Description Recommended Actions
No Gas Flow
Sample gas pump (option)
switched off
Switch on sample gas pump
(
page 6-5)
Membrane of sample gas
pump defective
Replace sample pump mem-
brane
Sample gas pump defective Replace sample gas pump
Solenoid valves (option) not
opened / defective
External valves:
Check connection between
valves and digital outputs
All valves:
Check valve seat and re-
place if need be
Replace solenoid valves
For valve control via serial
interface or digital inputs:
Any valve activated?
Gas path(s) polluted
Check gas path and sample
handling system for pollution
Clean gas path