2.2.1 Model-Specic Technical Data: X-STREAM XEGK
Temperatures
operational, max.
*)
0 … +50 °C / 32 … 122 °F
storage -20 … +70 °C / -4 … -158 °F
Weight, max 8 … 12 kg / 17.6 … 26.5 lb
IP or Type rating
IP 20 for indoor use,
protected against dripping water and direct sun light
Measurement channels, max. 3
Gas connections
max number 8
max for purging
(incl. / separate)
2 incl.
material PVDF; stainless steel (opt.)
sizes 6/4 mm;
1
4"
Power supply unit external; alternatively: internal wide range P/S
Power supply
Mains supply voltage uctuations are not to exceed
± 10 percent of the nominal voltage
nominal voltage DC 24 V 100–240 V
50 / 60 Hz
voltage range DC 10–30 V 85264 V
4763 Hz
nominal input current, max 2.5 A 1.3–0.7 A
Power input fuses
AC 230 V / T 3.15 A
5x20 mm (1 pcs)
AC 230 V / T 4 A
5x20 mm (2 pcs)
Electrical in- and outputs
power
3-pin XLR connector
IEC connector with integrated
power switch & fuse holders
signals
signal cables are connected using submin-D plugs or so-
ckets on the unit’s rear panel
special Ethernet: RJ45 socket; USB connectors
*)
: Limitations apply to selected measurement principles and ranges,
Measurement specications!