Cubemass DCI MODBUS RS485 Troubleshooting
Endress+Hauser 69
12 SENSOR
SW DAT
032 S: SENSOR SW DAT
$: # 032
Sensor DAT:
Error accessing the calibration values
stored in the S-DAT.
1. Check whether the S-DAT is correctly plugged
into the amplifier board.
2. Replace the S-DAT if it is defective.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
Spare part set number
Hardware revision code
3. Replace measuring electronics boards if
necessary.
13 TRANSM.
HW-DAT
041 S: TRANSM. HW DAT
$: # 041
Transmitter DAT:
1. T-DAT is defective.
2. T-DAT is not plugged into the
amplifier board or is missing.
1. Replace the T-DAT.
Check the spare part set number to ensure that
the new, replacement DAT is compatible with
the measuring electronics.
2. Plug the T-DAT into the amplifier board.
14 TRANSM.
SW-DAT
042 S: TRANSM. SW DAT
$: # 042
Transmitter DAT:
Error accessing the adjustment values
stored in the T-DAT.
1. Check whether the T-DAT is correctly plugged
into the amplifier board.
2. Replace the T-DAT if it is defective.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
Spare part set number
Hardware revision code
3. Replace measuring electronics boards if
necessary.
No. # 1xx Software error
143 A/C SW
COMPATIB.
121 S: A / C COMPATIB.
!: # 121
Due to different software versions, I/
O board and amplifier board are only
partially compatible (possibly
restricted functionality).
!
Note!
This message is only listed in the
error history.
Nothing is displayed on the
display.
Module with lower software version has either to be
updated by FieldCare with the required software
version or the module has to be replaced.
No. # 2xx Error in DAT / no communication
22 LOAD T-DAT 205 S: LOAD T-DAT
!: # 205
Transmitter DAT:
Data backup (downloading) to T-DAT
failed, or error when accessing
(uploading) the calibration values
stored in the T-DAT.
1. Check whether the T-DAT is correctly plugged
into the amplifier board.
2. Replace the T-DAT if it is defective.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
Spare part set number
Hardware revision code
3. Replace measuring electronics boards if
necessary.
23 SAVE T-DAT 206 S: SAVE T-DAT
!: # 206
27 COMMUNIC.
SENS
251 S: COMMUNICATION
I/O
$: # 251
Internal communication fault on the
amplifier board.
Replace the amplifier board.
28 COMMUNIC.
I/O
261 S: COMMUNICATION
I/O
$: # 261
No data reception between amplifier
and I/O board or faulty internal data
transfer.
Check the BUS contacts.
MODBUS No. Device status
message
(local display)
Cause Remedy / spare part
Register:
6859
Data type:
Integer
Register:
6821
Data type:
String (18 byte)