Cubemass DCI MODBUS RS485 Troubleshooting
Endress+Hauser 71
52 to 53 LOW FREQ. LIM. 379 S: LOW FREQ. LIM
$: # 379
The measuring tube oscillation
frequency is outside the permitted
range.
Causes:
–Measuring tube damaged
– Sensor defective or damaged
Contact your Endress+Hauser service organization.
53 UPPER FREQ. LIM. 380 S: UPPER FREQ. LIM
$: # 380
54 FLUIDTEMP. MIN. 381 S: FLUIDTEMP.MIN.
$: # 381
The temperature sensor on the
measuring tube is likely defective.
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
– Verify that the sensor signal cable connector is
correctly plugged into the amplifier board.
– Remote version:
Check sensor and transmitter terminal
connections No. 9 and 10.
55 FLUIDTEMP.
MAX.
382 S: FLUIDTEMP.MAX.
$: # 382
56 CARR.TEMP. MIN. 383 S: CARR.TEMP.MIN
$: # 383
The temperature sensor on the carrier
tube is likely defective.
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
– Verify that the sensor signal cable connector is
correctly plugged into the amplifier board.
– Remote version:
Check sensor and transmitter terminal
connections No. 11 and 12.
57 CARR.TEMP.
MAX.
384 S: CARR.TEMP.MAX
$: # 384
58 INL. SENS DEF 385 S: INL.SENS.DEF.
$: # 385
One of the measuring tube exciter
coils (inlet) is likely defective.
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
– Verify that the sensor signal cable connector is
correctly plugged into the amplifier board.
– Remote version:
Check sensor and transmitter terminal
connections No. 4, 5, 6 and 7.
59 OUTL. SENS. DEF 386 S: OUTL.SENS.DEF.
$: # 386
One of the measuring tube exciter
coils (outlet) is likely defective.
60 SEN. ASY.
EXCEED
387 S: SEN.ASY.EXCEED
$: # 387
One of the measuring tube exciter
coils is likely defective.
61 to 62 AMP. FAULT CH2
AMP. FAULT CH3
388
to
390
S: AMP. FAULT
$: # 388…390
Amplifier error Contact your Endress+Hauser service organization.
No. # 5xx → Application error
72 SW-DOWNLOAD 501 S: SW.-UPDATE ACT.
!: # 501
New amplifier or communication (I/
O module) software version is loaded.
Currently no other functions are
possible.
Wait until process is finished.
The device will restart automatically.
73 DOWN-UPLOAD
ACTIVE
502 S: UP-/DOWNLOAD
ACT.
!: # 502
Up- or downloading the device data
via configuration program. Currently
no other functions are possible.
Wait until process is finished.
76 OSC.AMP.LIM 586 S: OSC. AMP. LIMIT
$: # 586
The fluid properties do not allow a
continuation of the measurement.
Causes:
– Extremely high viscosity
– Process fluid is very
inhomogeneous (gas or solid
content)
Change or improve process conditions.
77 TUBE NOT OSC. 587 S: TUBE OSC. NOT
$: # 587
Extreme process conditions exist. The
measuring system can therefore not
be started.
Change or improve process conditions.
MODBUS No. Device status
message
(local display)
Cause Remedy / spare part
Register:
6859
Data type:
Integer
Register:
6821
Data type:
String (18 byte)