Proline Promass 83 PROFIBUS DP/PA Commissioning
Endress+Hauser 111
Example 2:
a0002682
Fig. 49: In this configuration example, modules that are not needed are replaced by the EMPTY_MODULE. The
Promass 83 GSD file is used.
With this configuration, the Analog Input function block 1(slot 1), the totalizer value TOTAL (slot
7) and the cyclic control of device functions CONTROL_BLOCK (slot 11) are activated. The mass
flow is read out cyclically by the measuring device by means of the Analog Input function block 1.
The totalizer is configured “without configuration”. In other words, in this example it only returns
the totalizer value for the mass flow by means of the TOTAL module and cannot be controlled by
the automation device.
Slot
sequence
Module
Byte length
input data
Byte length
output data
Description
1AI 5Analog Input function block 1
Output variable Mass flow
Data structure Page 103
2EMPTY_MODULE Empty
3EMPTY_MODULE Empty
4EMPTY_MODULE Empty
5EMPTY_MODULE Empty
6EMPTY_MODULE Empty
7TOTAL 5
Totalizer function block 1
Output variable for TOTAL Mass flow (factory setting)
8EMPTY_MODULE Empty
9EMPTY_MODULE Empty
10 EMPTY_MODULE Empty
11 CONTROL_BLOCK 1 Control of device functions