EasyManua.ls Logo

Endress+Hauser Proline Promass 83 - Diagnosis Transducer Block parameters

Endress+Hauser Proline Promass 83
178 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass 83 PROFIBUS DP/PA Trouble-shooting
Endress+Hauser 131
691 S SIM. FAILSAFE
! # 691
0x48
0x49
0x4A
UNCERTAIN Substitute set
(Substitute set
of failsafe
status)
O.K.
Low
High
Simulation
failsafe active
Cause
System error
Simulation of response to error is active.
Remedy:
Switch off simulation:
Access:
SUPERVISION SYSTEM OPERATION
SIM. FAILSAFE MODE ( OFF)
692 S SIM. MEASURAND
! # 692
0x60
0x61
0x62
UNCERTAIN Simulated
Value
(manually
specified
value)
O.K.
Low
High
Simulation
measured value
active
System error
Cause
Simulation is active
Remedy:
Switch off simulation:
Access:
SUPERVISION SYSTEM OPERATION
SIM. MEASURAND ( OFF)
698 S DEV. TEST AKT.
! # 698
0x60
0x61
0x62
UNCERTAIN Simulated
Value
(manually
specified
value)
O.K.
Low
High
Device test via
Fieldcheck active
Cause
System error
The measuring device is being checked on site via
the test and simulation device.
Remedy:
800 S: M. FL. DEV. LIMIT
!: # 800
0x40
0x41
0x42
UNCERTAIN Non Specific
(uncertain
status)
O.K.
Low
High
Mass Flow
outside the limit
Advanced Diagnostics:
The mass flow is outside the limit value, set in the
corresponding diagnosis function.
801 S: DENS. DEV. LIMIT
!: # 801
0x40
0x41
0x42
UNCERTAIN Non Specific
(uncertain
status)
O.K.
Low
High
Density outside
the limit
Advanced Diagnostics:
The density is outside the limit value, set in the
corresponding diagnosis function.
802 S: REF. D. DEV. LIM.
!: # 802
0x40
0x41
0x42
UNCERTAIN Non Specific
(uncertain
status)
O.K.
Low
High
Ref.Dens. outside
the limit
Advanced Diagnostics:
The reference density is outside the limit value, set
in the corresponding diagnosis function.
803 S: TEMP. DEV. LIMIT
!: # 803
0x40
0x41
0x42
UNCERTAIN Non Specific
(uncertain
status)
O.K.
Low
High
Temperature out.
the limit
Advanced Diagnostics:
The temperature is outside the limit value, set in the
corresponding diagnosis function.
804 S: T. DAMP. DEV. LIM.
!: # 804
0x40
0x41
0x42
UNCERTAIN Non Specific
(uncertain
status)
O.K.
Low
High
Tupe Damping
out. the limit
Advanced Diagnostics:
The tube damping is outside the limit value, set in
the corresponding diagnosis function.
805 S: E. D. SEN. DEV. LIM.
!: # 805
0x40
0x41
0x42
UNCERTAIN Non Specific
(uncertain
status)
O.K.
Low
High
Elec.-
mag.sens.out the
limit
Advanced Diagnostics:
The electrodynamic sensor is outside the limit value,
set in the corresponding diagnosis function.
No. Device status message
(local display)
PROFIBUS measured value status
Adv. diagnosis
message in
PROFIBUS
master
Cause/remedy
Quality code (HEX)
Quality status
Quality substatus
Limits

Table of Contents

Other manuals for Endress+Hauser Proline Promass 83

Related product manuals