Proline Promass 83 PROFIBUS DP/PA Technical data
Endress+Hauser 153
Zero point stability (Promass H, I):
Zero point stability (Promass E):
Sample calculation
F06-83xxxxxx-05-xx-xx-xx-000
Fig. 57: Maximum measured error in % of reading (example: Promass 83 F / DN 25)
Calculation example (mass flow, liquid):
Given: Promass 83 F / DN 25, flow measured value = 8000 kg/h
Max. measured error: ±0.10% ± [(zero point stability / measured value) x 100]% o.r.
Maximum measured error ±0.10% ±0.54 kg/h ÷ 8000 kg/h 100% = ±0.107%
DN Maximum full scale
value
in [kg/h] or [l/h]
Zero point stability
Promass H
in [kg/h] or [l/h]
Promass I
in [kg/h] or [l/h]
8 2000 0.20 0.20
15 6500 0.65 0.65
15 * 18000 1.8
25 18000 1.8 1.8
25 * 45000 4.5
40 45000 4.5 4.5
40 * 70000 7.0
50 70000 7.0 7.0
* DN 15, 25, 40 “FB” = Full bore versions of Promass I
DN Maximum full scale
value
[kg/h] or [l/h]
Zero point stability
[kg/h] or [l/h]
8 2000 0.200
15 6500 0.650
25 18000 1.80
40 45000 4.50
50 70000 7.00
0
±0.5
±0.2
±1.0
[%]
0 1 2 4 6 8 10 12 14 16 18
t/h