7 Maintenance and servicing
70 / 84
80116-938_V24392
2023-03
NE monitoring of asymmetry Testing and measuring equipment
Presets on the test specimen
Preferred return electrode type "split" (modify as appropriate under
<Protected settings>)
Electrode handle, monopolar – softCOAG – Effect 1
Erbe Art. No. Description
20192-127 Patient cable AE
or
20192-110 Patient cable AE, international
20190-045 Electrode handle ICC/ACC
Laboratory measuring cable
20194-070 Patient cable NE
or
20194-075 Patient cable NE, international
20100-033 Adapter cable NE
20100-101 VIO Testbox Symmetry/Resistance (NE
asymmetry/critical resistance)