13. Action-Tare-Memory (ATM)
13.1 Description of ATM Mode
ATM means action, then tare, then memory: -
a. Action = add or remove weight.
b. Tare = the above weight added on or removed will be tare off.
c. Memory = the above weight will be added to or deducted (in case of removal) from accumulated memory.
13.2 ATM Mode Settings
1. At Weighing mode, select the desired weight unit.
2. Go to ATM mode.
3. Refer to paragraph 10 and 11 if data related functions are needed.
4. Press [Set] then [Unit] to select Auto Accumulation target,
Gross = Gross weight will be accumulated.
Net = Net weight will be accumulated.
5. Press [Print/M+].
6. Instrument prompts for delay time (second). Delay time is the time interval (00 ~ 99 seconds): -
Between a valid stable weight result is obtained and before it is tare off and accumulated to memory.
Display time of total accumulated weight result (after all loadings are removed) and before it is
clear from print out memory.
7. Input delay time and press [Print/M+] to enter.
8. At this point, ATM mode is ready for use.
13.3 Start Using ATM
23
1. If zero weight cannot be obtained when unloaded, press [Zero]. Zero Indicator lights up when
instrument is at zero status.
2. Apply container. Instrument tares off the weight of the container.
3. Add or remove load. The weight result is displayed for the time interval set forth on step 6 of 13.2. Then
instrument will accumulate the weight result in memory then clear it from the display. Notes: -
Positive weight will be added to the accumulated memory.
Negative weight will be deducted from the accumulated memory.
4. Repeat step 3 until all weighing sequence is completed.
5. To recall the current total accumulated weight at any time, press [MR].
6. At this point: -
a. Press [Zero] to quit, or
b. Press [Print/M+] to print the current accumulated weight.
7. When all loads are removed, the total accumulated weight value will then be erased.
23
Weight changed less than 10d will not be processed.