14.1 Description of Animal Weighing Mode
Animal Weighing mode is used to weigh live animals.
14.2 Animal Weighing Mode Settings
1. At Weighing mode, select the desired weight unit.
2. Go to Animal Weighing mode.
3. Refer to paragraph 10 and 11 if data related functions are needed.
4. Press [Set] then [Unit] to select filter speed. 5 parameters (FLT 1 ~ 5) are available: -
• FLt 1 = Fastest speed with the lowest accuracy.
• FLt 3 = Normal speed. Recommended setting for most animal weighing applications.
• FLt 5 = Slowest speed with the highest accuracy.
5. Press [Print/M+] to save and to select weight release variation percentage. 10 parameters are
available (Off ~ 20%).
• rE oFF = auto-release disabled except all loads are removed.
• rE 2 = auto-release when weight varies ≥2% of rate capacity (or W1 for dual weighing
range/interval mode) or when all loads are removed.
• rE 20 = auto-release when weight varies ≥20% of rate capacity (or W1 for dual weighing
range/interval mode) or when all loads are removed.
6. Press [Unit] key to select the preferred weight release variation and press [Print/M+] to save.
7. Instrument is now ready for animal weighing application.
14.3 Start Using Animal Weighing
1. Get an animal on.
2. Instrument will calculate and hold the weight of an animal. The result being held flashes.
3. In case more animals have to be weighed in the same transaction, get other animals on. An updated
weight will be calculated, held and displayed