Operator‘s Manual FERITSCOPE
®
FMP30 Page 189
Malfunctions and Messages
Probe does not mea-
sure
Automatic
measurement acquisition
disabled.
Enable automatic mea-
surement acquisition
from the service func-
tions or trigger the ex-
ternal measurement
acquisition by pressing
the
key (or FINAL-
RES during the normal-
ization or calibration).
177
Probe defective. Use properly function-
ing probe.
--
Wrong readings Probe placed on speci-
men incorrectly (e.g.,
probe hovers over speci-
men).
Place probe on speci-
men correctly and ob-
serve the minimum lift-
off distance when lifting
off the probe.
34
Application selected that
has not bee set up for
this specimen.
Use APPL No and the
arrow keys to select the
correct Application.
40
Wrong base material
(correct measurements
with eddy current
probes are possible only
on non-ferromagnetic
base materials).
Use a specimen with a
non-ferromagnetic base
material.
--
Conducting metal close
the measurement loca-
tion when making mea-
surements using eddy
current probes.
(e.g., on the back side
of a pc-board).
Use a specimen with-
out a conducting metal.
--
Incorrect normalization
or calibration.
Perform a correct nor-
malization or calibra-
tion.
60
Malfunction/Message Cause Correction