Page 88 Operator‘s Manual FERITSCOPE
®
FMP30
Measuring
This measuring mode is advantageous if a mean values of the -ferrite con-
tent distribution is to be determined quickly within a reference area. This
mean value is added to a measurement series as a single reading.
With online output of the single readings via the interface port enabled,
only this mean value will be output as well.
The magnetic poles are subject to increased wear when moving the probe
across a surface.
Key sequ. /
Action
Detail of the display Explanation
Switch area measurement on from the menu
Service Functions/Measurement/Measur-
ing Mode ( 10.7.4 ‘Measuring Mode -
Standard/Area Measurement/Automatic
Measurement’, beginning on Page 177)
and place the probe on the specimen.
The displays the symbol for as long as
the probe is on the specimen. In addition,
an audible signal sounds when the probe is
placed on the specimen.
Move the probe with a slight and uniform
pressure across the specimen.
[Appl:]: Application number
[Fe%]: Displays
-ferrite contents in ferrite
percent (measured variable of the open Ap-
plication 10.7.6 ‘Measured Variable’,
beginning on Page 180)
[Blck:]: Number of the current block
[n=]: Number of the measurements / block
size stored in the Application
Lift the probe off the specimen.
An audible signal indicates the end of the
measurement.