Page 92 Operator‘s Manual FERITSCOPE
®
FMP30
Measuring
the closure of the block is indicated by a long audible signal (if the
measurement acquisition signal has been disabled, only the signal to
indicated the block closure will sound).
With a connected and powered up printer, it is now possible to print the
block result automatically after the block closure or upon ( 10.5.3 ‘Block
Result’, beginning on Page 162).
Key sequ. /
Action
Detail of the display Explanation
Make a measurement. The reading will ap-
pear on the display.
[Appl:]: Application number
[Fe%]: Displays
-ferrite contents in ferrite
percent (measured variable of the open Ap-
plication 10.7.6 ‘Measured Variable’,
beginning on Page 180)
[Blck:]: Number of the current block
[n=]: Number of the measurements / block
size stored in the Application
Perform the number of measurements re-
quired for an automatic block closure.
[]: Block is closed; no more measure-
ments can be taken in this particular block.
A new blocks starts automatically when the
next measurement is taken.
Individual readings / erroneous measurements can be overwritten in a
closed block ( 7.6.4 ‘Overwriting Individual Erroneous Measurements
at a Later Time’, beginning on Page 103).